• BACKGROUND
    • The feasibility of percutaneous closure ventricular septal defects (VSD) in children has been previously proven. However, data on long-term outcomes are limited. We aim to evaluate the long-term outcome of our experience with percutaneous closure of VSD using various occluders.
  • METHODS
    • Retrospective institutional analysis of children who underwent transcatheter closure of perimembranous and muscular VSDs between September 2012 and February 2020. Patient demographics, procedural, and long-term follow-up data were comprehensively analyzed. Patients who lost to follow-up within two years post-procedure were excluded.
  • RESULTS
    • We identified 75 patients (54.7% males) with a median of 66 months (IQR, 46-96). The closure success rate at one year was 95.7%. Complete heart block was detected in two patients early post-procedure and resolved with steroids. The VSDs were perimembranous (52%), muscular (33.33%), and residual (14.67%). Implanted devices were Pfm Nit-Occlud LeˆVSD Coil (42.7%), HyperionTM VSD Muscular Occluder (28%), Amplatzer VSD muscular occluder (10.7%), Amplatzer Duct Occluder (14.7%), Occlutech Muscular VSD Occluder (2.7%), and Amplatzer Duct Occluder II (1.3%). No new arrhythmia or valve regurgitation was detected after two years post-procedure. Persisted complications on long-term follow-up included: residual shunting in 3(4%), mild tricuspid regurgitation in 2(2.7%), and aortic regurgitation in 2(2.7%), with one immediate post-catheterization mild aortic regurgitation worsened during follow-up, requiring surgical repair of VSD three years after device implantation. No deaths were reported.
  • CONCLUSION
    • Long-term outcomes of pediatric transcatheter VSD closure using different devices are satisfactory. Post-procedural adverse events are limited, but long-term surveillance is necessary to monitor their progression.